Multiple intracranial aneurysms (MIA) mean two or more aneurysms, which are common clinically. Due to the increase in the number of aneurysms and diversity of their space distribution, MIA has been difficult in treating intracranial aneurysms. For multiple intracranial aneurysms, a safe and effective therapeutic regimen should be provided according to the patient’s condition and the characteristics of aneurysms[1-2]. With the development of interventional therapy concepts and materials, most aneurysms can be treated by endovascular embolization. The stent technology, balloon assisted technology and double catheter technology also make wide-neck aneurysms no longer become the difficulties of interventional therapy[3]. For the treatment of MIA, endovascular intervention may also be preferred. It is not affected by the site of aneurysms, avoiding pull to the brain tissue; in addition, it features small incision, fast recovery and significant superiority[4-5]. When the lesion distribution space involves bilateral and even supratentorial/subtentorial encephalocoele, all the aneurysms cannot be treated with unilateral craniotomy once. In this case, embolization can be used to treat only the responsible aneurysm, and then the aneurysm can be treated by stages according to the risk of bleeding[6].
Patient Information
• Male, 63Y, multiple carotid aneurysms.
Surgery plan
• Due to the patient’s complicated condition, embolization was performed by stages. Till today, two responsible aneurysms have undergone embolization. This time, we plan to embolize the third aneurysm with a stent together with embolism.
Pre-operative imaging
• 3D-DSA before operation showed that there were many aneurysms in the internal carotid artery segment, and the size of aneurysms embolized in this operation was 3.1 mm * 3.7 mm.

Procedure
• Guided by the approach graph during the operation, using the double catheter technique, put the stent microcatheter (TJWY’s Frepass®18 microcatheter) and microcatheter for embolism (TJWY’s Frepass®14 microcatheter) in place and then make them into a basket with TJWY’s Perdenser® 3mm*6cm coil system-3D. It can be seen the basket is securely attached to the wall with satisfactory effect.

• Then use one 2mm*6cm and one 2mm*4cm Perfiller® expandable coil system for embolism-2D to successively fill the aneurysm and finally close up with a 2mm*6cm Perdenser® coil system for embolism-2D.

After Operation
• After liberating the stent, perform radiography again to see that the aneurysm is densely embolized, there is no contrast agent residue in the cavity and the operation ends with an ideal effect.

In this case, Perdenser® Coil System for Embolism was used for making basket. Since Perdenser® Coil System for Embolism is preshaped coil, it can be used for stably making basket. With 360° Ω-shaped design, it is well conformal and the basket can perfectly attach to the wall in a secure manner. An excellent basket is not only helpful for aneurysm wall protection, but also provides a good foundation for subsequent filling. When Perfiller® expandable coil system for embolism is used for filling, its polymer expandable silk can form aquagel after expansion with water absorption. The aquagel fills in the first-level spiral uniformly and densely to embolize the aneurysm body and neck.
The latest marketed Perfiller® Expandable Coil System for Embolism (polymer thread) is divided into complex type (3D) and spiral type (2D). The polymer thread expands in water, which can effectively increase the embolism density, so as to achieve the effect of embolization density. Since the expanded polymer thread may effectively enhance the coil stability and avoid coil system compression due to blood flow impact, the aneurysm neck can be protected for a long time and “dog ear sign” can be avoided and long term recurrence rate can be reduced effectively.

When the polymer thread swells when exposed to water, it is smaller than the first-level spiral diameter of the spring coil. The use process is not limited by time, and it will not squeeze the adjacent spring coil and aneurysm wall. It will only expand to the place with gaps, and there will be no case of aneurysm rupture caused by over-expansion.
With the brand-new push system, the mechanical transmission of the push rod is more accurate, and the push process is easier and smoother.

TJWY’s independently developed Perdenser® Coil System for Embolism is divided into complex type and spiral type. The complex type can be used to make basket with its 360° Ω-shaped design with freely turning point to effectively protect the aneurysm wall and avoid its rupture. An excellent basket performance is not only helpful for aneurysm wall protection, but also provides a good foundation for subsequent filling. The head end of the coil system is designed with a polymer ball cap, which can ensure no damage to the aneurysm wall and avoid breaking the aneurysm during the operation.
Excellent gap detection and filling ability effectively ensures that dense embolism can be achieved during surgery. Hundreds of models are available for basket making to packing and close-up.
According to the different diameters of the coils, the system can meet the filling requirements of aneurysms of different sizes and support the whole process operation from basket forming to packing and close-up. Taijie Weiye’s utrasoft coil system for embolism has good flexibility, which can be used for basket forming and packing in the micoaneurysms treatment, but also can be used as close-up coil in the treatment of large-and medium-sized aneurysms, providing safe and reliable performance.
As the first fast-detachment coil system in China, the detachment method is stable and fast, which can effectively reduce the operation time. Especially in the case of aneurysm rupture treatment, it can race against the clock to reduce the patient's risk.
References:
[1] Qin Shangzhen, Ma Lianting, Gong Jie, et al. Diagnosis and treatment of multiple intracranial aneurysms [J]. Chinese Journal of Neurosurgery, 2012, 28(9): 872-4
[2]SibiliaJ, Gottenberg J E, Mariette X. Rituximab: a new therapeutic alternativein rheumatoid arthritis[J]. Joint Bone Spine, 2008, 75(5):526-32.
[3]CoiffierB, Lepage E, Briere J, et al. CHOP chemotherapy plus rituximab compared withCHOP alone in elderly patients with diffuse large- B- cell lymphoma[J]. N Engl J Med, 2002, 346(4):235-42.
[4]KishiJ, Nanki T, Watanabe K, et al. A case of rituximab- induced interstitialpneumonitis observed in systemic lupus erythematosus[J]. Rheumatology (Oxford), 2009, 48(4): 447-8.
[5]KawabataY,Nakazawa T, Fukuda S. Endovascular embolization of branch-incorporatedcerebralaneurysms[J]. Neuroradiol J, 2017, 30(6):600-606.[6]YangY, Su W, Meng Q. Endovascular treatment of ruptured true posterior communicating artery aneurysms[J]. Turk Neurosurg, 2015, 25(1):73-77.